We provide best solution for continuous monitoring in various forms of mercury. Our Hg mercury analyzer applies latest and most accurate Cold Vapour Atomic Fluorescence Technology, integrated with dilution sampling system to reduce moisture, heat and other background interferences.
By combining Catalytic Pyrolysis Elemental Conversion Technology and Atomic Fluorescence Analytical Technology, We offers solution on Elemental, Ionic and Total Mercury Monitoring.
Principle | Cold Vapor Atomic Fluorescence (CVAFS) |
Range | 0~20, 0~200, 0~2000 μg/m3 (Selective) |
Lower Detectable Limit | 1 ng/m3 (40:1 dilution ratio) |
Linearity | ≤±0.5%F.S. |
Repeatability | ≤±1%F.S. |
Zero Drift | ≤±0.5%F.S./24h |
Span Drift | ≤±0.5%F.S./24h |
Response Time | 30s |
Operational Temperature | -20~50 °C |
Consists of dilution probe, heated sampling line, elemental converter, mercury analyzer, and calibration unit.
The Performs online continuous measurement on mercury by adopting atomic fluorescence technology, without any pre-concentrating sample on gold-trap, the system sharply reduces the maintenance cost. Unique optical design offers low interference and high liability.
Dilution probe adopts SS316 provides durable and accurate sampling.
PTFE sampling line with variable temperature control between 120 and 200 °C as well as adjustable length.
Elemental mercury span generation is achieved by precise temperature and flow control with permeation technology.
The Pyrolysis Converter offers more than 95% conversion rate which enables simultaneous measurement on elemental, Ionic and Total Mercury.