• Green Enviro International Pvt Ltd
ISO 9001 : 2015     ISO 14001 : 2015     ISO 45001 : 2018    
Mercury Analyzer

We provide best solution for continuous monitoring in various forms of mercury. Our Hg mercury analyzer applies latest and most accurate Cold Vapour Atomic Fluorescence Technology, integrated with dilution sampling system to reduce moisture, heat and other background interferences.

Principle

By combining Catalytic Pyrolysis Elemental Conversion Technology and Atomic Fluorescence Analytical Technology, We offers solution on Elemental, Ionic and Total Mercury Monitoring.

Feature and Benefit

  • Accurate measuring elemental, ionic and total mercury.
  • High sensitivity achieved by reducing moisture and background interference based on high dilution ratio.
  • Real time online monitoring without high expense on batch processing.
  • No consumables nor hazards produced during process, low maintenance.
  • Specification

    Principle Cold Vapor Atomic Fluorescence (CVAFS)
    Range 0~20, 0~200, 0~2000 μg/m3 (Selective)
    Lower Detectable Limit 1 ng/m3 (40:1 dilution ratio)
    Linearity ≤±0.5%F.S.
    Repeatability ≤±1%F.S.
    Zero Drift ≤±0.5%F.S./24h
    Span Drift ≤±0.5%F.S./24h
    Response Time 30s
    Operational Temperature -20~50 °C

    Components

    Consists of dilution probe, heated sampling line, elemental converter, mercury analyzer, and calibration unit.

    Mercury Analyzer

    The Performs online continuous measurement on mercury by adopting atomic fluorescence technology, without any pre-concentrating sample on gold-trap, the system sharply reduces the maintenance cost. Unique optical design offers low interference and high liability.

    Dilution probe

    Dilution probe adopts SS316 provides durable and accurate sampling.

    Heated sampling line

    PTFE sampling line with variable temperature control between 120 and 200 °C as well as adjustable length.

    Calibrator

    Elemental mercury span generation is achieved by precise temperature and flow control with permeation technology.

    Converter

    The Pyrolysis Converter offers more than 95% conversion rate which enables simultaneous measurement on elemental, Ionic and Total Mercury.